There are several modes for positioning series windows in the study window. The selected mode is saved when the DICOM Viewer is closed.
To maximize any window of the series and restore its size, proceed as follows:
double-click with the left mouse button on the series window;
double-click with the left mouse button on the title of the series window;
click on the Expand button in the right-hand upper corner of the series window.
If some tool, such as the tool for measurements or render annotations, has been activated with the left mouse button, you can expand a window or restore its size in two ways:
double-click with the left mouse button on the title of the series window;
click on the Expand button in the right-hand upper corner of the series window.
This mode is active by default. If two or more series are open, they will be located in the window and scaled automatically, occupying the entire window. Three series are displayed in Fig. 2.9.
To activate this mode, click the button or select the Automatic series arrangement
item from the image context menu.
This mode allows you to open only one series in the study window. If you try to open another series, the current series will be closed. In this case, the DICOM Viewer will request confirmation for closing the current series.
To activate this mode, click the button or select the Stacked series arrangement item
from the image context menu.
In this mode the study window is split into parts, and a series window can be opened in each of them.
A grid may have from 1 to 100 cells. The maximum number of rows and columns is 10.
Fig. 2.10 illustrates a 2*4 grid with three cells filled.
To set up a 1x2, 2x2 or 2x3 grid, use the ,
,
buttons on the toolbar. These
grid configuration types are available from the image context menu.
If the selected grid configuration has fewer cells than there are open series in the study, a warning dialog box pops up (Fig. 2.11), in which the program asks for a permission to close the open series.
Click OK to close the series, or Cancel to cancel.
There are two ways to disable the warning dialog box:
Check the Don’t show this dialog again box in the dialog (Fig. 2.11) and click the OK button.
In the main menu, select Options->Settings...->Modules->Image Viewer, then choose the Image arrangement tab and uncheck the Confirm series closure on the arrangement changing box. For more information on the module settings see Section 15.5.1.
Click OK to apply the settings or Cancel to cancel.
To define your own grid configuration, click the button or select the Custom series
arrangement item from the image context menu, and set up the required parameters. This
configuration can be set by default (see Section 15.5.1).
For customized grids, the DICOM Viewer allows for auto-fill of the cells with the current study series. The auto-fill function is disabled by default. To enable it, in the main menu, select Options->Settings...->Modules->Image Viewer, then choose the Image arrangement tab and check the Turn on auto-fill in series box. When a series is opened in a new tab, the other series are opened in the cells available. The cells are sorted by serial number.
When positioning series on the grid, the DICOM Viewer allows you to move the boundary
between cells. To do this, move the cursor to the panel border so that it would look like or
and drag the border holding the left mouse button. Setting the priority of horizontal or vertical
separator is described in Section 15.5.1.