2.19 Measurements

PIC To avoid errors caused by the discrepancy between the image size and the real size, Calibration tool should be used when working with projection images (e.g. X-ray images) (see Section 2.36).
PIC The measurement accuracy is up to one screen pixel. As a screen pixel is smaller than a source image pixel, the true linear measurement accuracy is up to one source image pixel. Errors may also occur in density measurements. The most accurate measurements are performed with bilinear interpolation (see Section 2.5).

The following tools are used to measure various parameters: Ruler, Polygonal ruler, Angle, Cobb angle, Point value, ROI rectangle, ROI ellipse, ROI polygon. To select one of these tools, click on the arrow on the right side of the tool selection button. The button will look different depending on the selected tool:

PIC Ruler (selected by default)
PIC Polygonal ruler
PIC Angle
PIC Cobb angle
PIC Point value
PIC ROI rectangle
PIC ROI ellipse
PIC ROI polygon
PIC Cardiothoracic ratio(PRO)

To activate or deactivate the tool that is currently selected, just click on the tool selection button. If some tool is activated, the button is highlighted. The drawn objects will be displayed in the window while it is open, and you can pan, zoom or rotate them together with the image.

If the series is open in multiple windows or tabs, the measurements made in any window, appear in all the other windows.

2.19.1 Ruler

To measure distance in an image:

  1. Activate the Ruler PIC tool on the toolbar by clicking the left/right/middle mouse button. To continue work with this tool, use the button with which the tool was activated. To learn more about tool control, see Section 1.14.

  2. You can measure the distance in two ways:

  3. To cancel an incomplete measurement, press Esc.

This tool is also available from the image context menu and from the Image main menu item. Activation is only possible with the left mouse button.

The measurement results are not saved in the project file. To save the measurement results, export the current image to a file (see Section 2.27) or use marker lines (see Section 5.12.3).

2.19.2 Displaying angles between intersecting rulers

____________________________________________________________________________________________

Functionality is available in the Pro edition
____________________________________________________________________________________________

The DICOM Viewer provides an opportunity to see the base and the adjacent angle when two Ruler  PIC tools cut each other in the Image Viewer, MPR reconstruction, Vessel analysis, Coronary artery analysis, PET analysis, and Cardiac function analysis tabs.

To see the angle between two intersecting Ruler tools, check the Show angles between rulers(PRO) box in the Tool options dialog box. The box is checked by default. You can choose whether only the base angle or both (the base and the adjacent) angles are to be shown (see Section 2.19.9).

The base angle between the rulers is marked with a single arc, and the adjacent angle is marked with a double arc. The angle value is shown in degrees as a separate note.

2.19.3 Polygonal Ruler

To perform plygonal linear measurements:

  1. Activate the Polygonal ruler PIC tool by clicking the left/right/middle mouse button. To continue work with this tool, use the button with which the tool was activated. To learn more about tool control, see Section 1.14.

  2. Mark the first point on the toolbar by clicking the mouse button.

  3. Drag the cursor over the screen. The distance from the first point to the current point will be displayed beside the line.

  4. To fix the current point, click the mouse button.

  5. Repeat Steps 3 and 4 until the last but one point is fixed.

  6. Fix the last point by double-clicking the mouse.

  7. To cancel an incomplete measurement, press Esc.

To move the linear measurement point:

  1. Locate the cursor on the point to be moved.

  2. Drag the point, holding down the left mouse button or the button to which this tool is assigned;

  3. Release the mouse button.

To delete a point:

  1. Locate the cursor on the point to be deleted.

  2. Right-click the mouse and select the Remove point command.

To insert a point:

  1. Locate the cursor on the ruler where you want to insert a point.

  2. Right-click the mouse and select the Insert point command.

This tool is also available from the image context menu and from the Image main menu item. Activation is only possible with the left mouse button.

The measurement results are not saved in the project file. To save the measurement results, export the current image to a file (see Section 2.27) or use polygonal marker lines (see Section 5.12.4).

2.19.4 Angle

To measure an angle:

  1. Activate the Angle PIC tool from the toolbar by clicking the left/right/middle mouse button. To continue work with this tool, use the button with which the tool was activated. To learn more about tool control, see Section 1.14.

  2. Put two points on the image, clicking the mouse button. The second point is the angle apex.

  3. Move the cursor over the screen to set the other side of the angle.

  4. Click the mouse button to fix the second side of the angle.

  5. To cancel an incomplete measurement, press Esc.

The DICOM Viewer allows you to build as many angles as you need. To deactivate the tool, click on the tool selection button.

This tool is also available from the image context menu and from the Image main menu item. Activation is only possible with the left mouse button.

The measurement results are not saved in the project file. To save the measurement results, export the current image to a file (see Section 2.27) or use polygonal marker lines (see Section 5.12.4).

2.19.5 Cobb angle meter

To measure a Cobb angle:

  1. Activate the Cobb angle PIC tool from the toolbar by clicking the left/right/middle mouse button. To continue work with this tool, use the button with which the tool was activated. To learn more about tool control, see Section 1.14.

  2. Draw a line that runs along the border of one of the vertebrae. To do this, click the mouse button to set two points through which the line must pass.

  3. Likewise, construct a line for the second vertebra.

  4. To cancel an incomplete measurement, press Esc.

To display additional parameters, hover the mouse over the measurement tool or the measurement value. Right-click the mouse and select the Tool options... item from the context menu.

In the Tool options dialog box, provide the measurement name, as well as the color, the line width, and the font size for the angle value. To join the measurement value to the angle with a dashed line, check the Footnote line box.

To display the value of the adjacent angle, check the Adjacent angle box.

To show the angle value at the intersection of segments, check the Show intersection box. If this function is disabled or the segments do not intersect, the angle value is shown only at the intersection of normals. In Fig. 2.29 the option Show intersection is disabled, and in Fig. 2.30 is enabled.


PIC

Figure 2.29: The Show intersection option is disabled



PIC

Figure 2.30: The Show intersection option is enabled


To use these parameters by default, check the box Set as default.

To apply the settings, click OK. To cancel, click CANCEL.

To delete mesurement, locate the cursor on it, right-click the mouse and select the Remove object. For more information see Section 2.19.12.

This tool is also available from the image context menu and from the Image main menu item. Activation is only possible with the left mouse button.

The measurement results are not saved in the project file. To save the measurement results, export the current image to a file (see Section 2.27) or use polygonal marker lines (see Section 5.12.4).

2.19.6 Measure Intensity at a Point

PIC If non-bilinear interpolation filters are used, significant errors may occur when intensity is measured in a certain point. In this case, the intensity value will be marked by an asterisk. For more accurate measurements, use the bilinear interpolation filter (see Section 2.5).

To measure intensity at some point:

  1. Activate the Point value PIC tool from the toolbar by clicking the left/right/middle mouse button. To continue work with this tool, use the button with which the tool was activated. To learn more about tool control, see Section 1.14.

  2. Click the mouse button on the target point.

The point will be highlighted at the image, and the intensity value will be displayed next to it. It is shown in the "k HU" format for original images and in "k", where k is the intensity value at the point. If the value is inaccurate, you will see a "*" symbol next to it.

If you apply a filter to an image, you will see an asterisk "*" symbol next to the intensity value. It informs the user that the measurement results are affected by the filter applied (see Section 2.6).

PIC An image to which a filter has been applied is not original, and the measurement results may be inaccurate.

To adjust the Point value tool, locate the cursor on the point or the mesurement results on the image. Right-click the mouse and select the Tool options... item from the context menu. In the Tool options dialog box enter the measurement name. Set the color and the font size. To connect the block with the measurement results with a dotted line, the Footnote line box must be checked. To use these parameters by default, check the box Set as default.

To apply the settings, click OK. To cancel, click CANCEL.

To delete mesurement, locate the cursor on it, right-click the mouse and select the Remove object.

This tool is also available from the image context menu and from the Image main menu item. Activation is only possible with the left mouse button.

The measurement results are not saved in the project file. To save the measurement results, export the current image to a file (see Section 2.27).

2.19.7 Measure the Intensity Average and Standard Deviation in an Area

To measure the average intensity value and standard deviation in a particular area:

  1. Activate one of the ROI ... tools (rectangle PIC , ellipse PIC , or polygon PIC ) from the toolbar by clicking the left/right/middle mouse button. To continue work with this tool, use the button with which the tool was activated. To learn more about tool control, see Section 1.14.

  2. You can draw a rectangle in two ways:

  3. To draw an ellipse, just draw the rectangle the ellipse is inscribed in.

  4. To draw a polygon, click the mouse button to mark each apex and then double-click the mouse to finish.

  5. To cancel an incomplete measurement, press Esc.

These tools are also available from the image context menu and from the Image main menu item. Activation is only possible with the left mouse button.

The following parameters will be displayed next to the highlighted area:

If you apply a filter to an image, you will see an asterisk "*" symbol in the ROI measurement results line. It informs the user that the measurement results are affected by the filter applied (see Section 2.6).

PIC An image to which a filter has been applied is not original, and the measurement results may be inaccurate.

2.19.8 Measuring the Cardiothoracic Ratio (CTR)

____________________________________________________________________________________________

Functionality is available in the Pro edition
____________________________________________________________________________________________

To measure the cardiothoracic ratio, proceed as follows:

  1. Activate the Cardiothoracic ratio PIC tool from the toolbar by clicking the left/right/middle mouse button. To continue work with this tool, use the button with which the tool was activated. To learn more about tool control, see Section 1.14.

  2. Measure the thoracic diameter by marking points on the inside of the ninth rib (points 1 and 2 in Fig. 2.31).

  3. Build the vertebral column axis (points 3 and 4 in Fig. 2.31).

  4. Measure the transverse cardiac diameter by marking the outermost points of the left and the right contours (points 5 and 6 in Fig. 2.31).

  5. To cancel an incomplete measurement, press Esc.

Next to the measurement, you can see the following results:

This tool is also available from the image context menu and from the Image main menu item. Activation is only possible with the left mouse button.


PIC

Figure 2.31: Measuring the cardiothoracic ratio


2.19.9 Drawing Parameters

To set the default drawing parameters for measurement tools:

  1. Click on the arrow on the right side of the measurement tool selection button.

  2. Select the item Default tool options....

  3. In the dialog box that appears, set the color by clicking on the color box.

  4. Set the line thickness and the font size.

  5. Set length and area units (millimetres or centimetres).

  6. To connect the footnote and the measurement of the dotted line, check the box Footnote line.

  7. To see the angle between two intersecting Ruler tools, check the Show angles between rulers(PRO)box. The box is checked by default.

  8. Choose the angle display option:

  9. To apply the settings, click OK. To cancel, click CANCEL.

Drawing parameters are available as well from the image context menu and from the Image main menu item.

The settings will be applied only to the measurements conducted in the flat image view window. For multiplanar reconstruction, similar settings should be saved separately.

To edit the drawing parameters of an existing measurement:

  1. Locate the cursor on the measurement to highlight a line or a point.

  2. Right-click the mouse.

  3. Select the Tool options... item from the context menu.

  4. In the dialog box that appears, set the parameters in the same way as the default drawing parameters.

  5. To use these parameters by default, check the box Set as default. Set the measurement name if necessary.

2.19.10 Move Measurements

To move a measurement:

  1. Locate the cursor on the measurement to highlight the line or point or to magnify the cross, marking the angular point. Do not locate the cursor on angle points.

  2. Move the measurement, holding any mouse button.

2.19.11 Move Measurement Results

To move a measurement result:

  1. Locate the cursor on the measurement so that it is highlighted by a dotted frame.

  2. Move the measurement result, holding down the left mouse button or the button to which this tool is assigned;

If the cursor is located on the measurement, the corresponding measurement result will be highlighted, and vice versa.

2.19.12 Delete Measurements

To delete annotations and measurements, locate the cursor on the measurement to highlight a line or a point and press the Delete key on the keyboard, or right-click the mouse and select the Remove object item from the context menu.

To delete all annotaions and measurements, click the Delete all annotations and measurements PIC button on the toolbar or select the corresponding item from the Image main menu. In the dialog box that appears, Click YES to delete or NO to cancel. If you want the DICOM Viewer to always delete annotations and measurements without this confirmation, check the box Don‘t show this dialog again and click YES. If you want the DICOM Viewer to show the confirmation dialog, check the box Ask before deleting all annotations and measurements (see Section 16.7.1).

PIC Please note that all other graphic objects created manually will be deleted as well.

2.19.13 Measuring Dimensions and Area during Ultrasound Imaging

The Ruler, Polygonal ruler, ROI rectangle, ROI ellipse and ROI polygon tools may be used for linear and area measurements on ultrasound images (series with US modality).

Ultrasound study series may contain data on the areas of images with a certain pixel size.

If measuring tools are applied to the area of the image for which the pixel size has been determined, then the measurement units provided for in the tool parameters will be used for linear and area measurements (see Section 2.19.9).

If a line or a part of the measuring area stretches beyond the area for which the pixel size has been determined, linear and area measurements will be performed in pixels.

If the series does not contain any data on the areas for which the pixel size has been determined, linear and area measurements will be performed in pixels.

2.19.14 Measuring the Time Interval and the Blood Flow Velocity during Ultrasound Imaging

The Ruler and Polygonal ruler tools may be used for measuring the time interval and the blood flow velocity when performing ultrasound examination (series with US modality) with dopplergram.

PIC For the Polygonal ruler tool, the values of the time interval and blood flow velocity are only shown when the measurement is made by two points.

When measurements are performed, the measurement value is shown on the dopplergram image next to the segment built, in the X: x s Y: y cm/s format, where x is the value of the time interval in seconds (s) accurate up to hundredth; y is the value of the blood flow velocity in cm/s, accurate up to hundredth.

If the measurement is fully or partly performed beyond the Doppler area, the linear dimension will be shown next to the segment built. In this case, the measurement units depend on the current algorithm and program settings.

If no image areas are suitable for evaluation of the time interval and the blood flow velocity, then when a measurement is made, the linear value will be shown next to the segment built. In this case, the measurement units depend on the current algorithm and program settings.